AirChek 2000 SKC Personal Sample Pump
Setting the Standards of a New Millennium
Programmable: 5 to 3250 ml/min Whether using the AirChek 2000 air sampling pump manually or connected to a PC, it's the easiest pump to operate. The versatile AirChek 2000 can be used for high flow sampling up to 3250 ml/min or for flow rates down to 5 ml/min with an easy-to-use low flow adapter kit.
The AirChek 2000 air sampling pump is ideal for industrial hygiene studies (including personal and area sampling), environmental testing, occupational health sampling and indoor air quality investigations. The AirChek 2000 air sampling pump is typically used with collecting devices such as: Sorbent sample tubes and Long-duration detector tubes for collecting most hazardous chemical gases and vapors from the air
Cyclones for collecting respirable dust
IOM samplers for collecting inhalable dust
Filters for collecting total dust, asbestos, or chemical hazards
Tedlar sample bags for collecting airborne chemical hazards
Impingers for collecting reactive chemicals into liquids
Advanced Electronic Flow Control
The AirChek 2000 air sampling pump with its patented internal isothermal flow sensor (patent 585,892,160) brings advanced electronic flow control to air sampling. This new air sampling technology allows the user to program the desired flow rate into the sample pump with an accuracy of 5% using the keypad or the optional DataTrac 2000 software. The internal isothermal flow sensor measures flow directly and acts as a secondary standard; once calibrated against an external primary standard, the flow sensor constantly maintains the flow rate. Built-in sensors automatically correct the flow rate for variations in temperature and pressure.
The AirChek 2000
Advanced Technology to Meet the Demands of the New Millennium PC-programmable with DataTrac 2000 software Program for start/stop times, intermittent and repeated sampling, and download sampling data to a PC for storage and analysis with optional DataTrac 2000 software.
Advanced electronic flow control
Internal isothermal flow sensor maintains flow setting.
Advanced technology Built-in sensors automatically correct the flow for variations in temperature and atmospheric pressure.
Wide sample pump flow range With its wide flow range of 5 to 3250 ml/min, the AirChek 2000 is the best choice for most air sampling applications; a low flow range of 5 to 500 ml/min requires an easy-to-use low flow adapter kit (low flow does not include some electronic readout options).
Compact, lightweight, and ultra-quiet For greater worker acceptance, the AirChek 2000 is lightweight (22 ounces) and portable. The Airchek 2000 air sampling pump is outfitted with a rugged impact-resistant RFI/EMI-shielded case with removable battery pack.
No tools required The simple 3-button keypad does it all.
Intrinsically-safe Versatile for all industries and safe in explosive environments; UL and cUL listed.
Continuous air sample volume calculations All air sampling parameters are continuously updated and available at the touch of a button.
Automatically corrects for temperature and atmospheric pressure Built-in sensors correct the flow rate for variations in temperature and atmospheric pressure.
RFI/EMI-shielded, water-resistant case Shielding protects the air flow from outside interference ensuring accurate air volumes and data validity; covers protect battery charging jack and computer interface port; CE-approved.
Security system protects data A security code, required to change the sample pump operating parameters, minimizes accidental changes and maintains sample validity.
Lithium backup battery Internal lithium backup battery preserves data when the battery pack is removed; factory calibration and program schedules are preserved.
Multi-tube sampling feature Optional multiple adjustable low flow holders allow simultaneous air sampling with 2, 3, or 4 tubes.
Real-time clock Optional display in 12-hour standard with a.m. and p.m. indicators or in 24-hour military time.
Automatic features maintain sample integrity: Auto shut-off with low battery or restricted flow.
Adjustable flow fault shutdown from 5 seconds to 4 minutes when used with a PC and DataTrac 2000 Software.
Auto-restart from flow fault attempted every 5 minutes for a maximum of 10 times (can be disabled with DataTrac 2000 software).
Run-time data stored in memory.
Easy Sample Pump Operation
Easy manual programming Set flow rate and run-time without screwdrivers. Simply use the 3-button keypad to control the AirChek 2000 air sampling pump manually. A convenient timed shutdown feature allows you to program the AirChek 2000 air sampler to run for a set time from 1 to 999 minutes; this is useful for STELs or repetitive sample runs.
Large easy-to-read LCD
The large easy-to-read LCD displays a direct readout of : air flow rate air volume run-time air temperature atmospheric pressure battery status time-of-day flow fault Data is continuously updated and can be displayed during air sampling by simply pressing a single button on the pump. The LCD shows a battery status icon constantly and a flashing flow fault icon during restricted or blocked flow. The flashing LED on the top of the sample pump indicates that the pump is running.
Flow fault features
Auto shut-off with low battery or drop in flow of 5% or more. Adjustable flow fault shutdown from 5 seconds to 4 minutes when used with a PC and DataTrac 2000 Software. Auto-restart from flow fault attempted every 5 minutes for a maximum of 10 times (can be changed with DataTrac 2000 software). Run-time data stored in memory.
AirChek 2000 Now with CalChek*
Direct Communication to a Primary Standard Calibrate the Airchek 2000 pump to the desired flow using the CalChek Communicator and a DC-Lite calibrator; no manual adjustment needed!
The CalChek automatic calibration feature is available when calibrating an Airchek 2000 pump (version 2.59 or higher) with a DC-Lite calibrator model 717-01 (10 ml/min to 12 L/min). A CalChek Communicator smart adapter is required for communication between the pump and the calibrator. Optional DataTrac 2000 software (version 3.59 or higher) can be used to expand the documentation capabilities of this system.
Fast, Easy, and Accurate Calibration CalChek allows automatic field calibration and reduces the need for frequent factory calibration.
Flexible Calibration System CalChek provides two calibration options: Single Point Calibration quickly verifies flow at a single point before and after sampling for accuracy.
Multiple Point Calibration corrects across a range of flows (750 to 3250 ml/min) in just minutes for maintenance and quality programs. * US Patent 6,227,031
The AirChek2000 Air Sampling Pump
Performance Profile Operating flow range 750 to 3250 ml/min 5 to 500 with low flow adapter kit Constant flow compensation (inches water back pressure) 3250 ml/min up to 15 inches 2500 ml/min up to 20 inches 2000 ml/min up to 30 inches 750 ml/min up to 40 inches Volumetric flow Held within 2% of the set flow from 1 to 3 liters and 4% below 1 liter Flow fault If flow drops by 5% or more, pump stops and holds historical data. Auto-restart is attempted every 5 minutes up to 10 times. Run-time 8 hours with the 2.0 Ah capacity battery pack
Accuracies Flow Atmospheric pressure Timing accuracy 3% of setpoint after calibration to desired flow 0.3 inches Hg 1 minute/month at 25 C Temperature range Operating Charging Storage
32 to 113 F (0 to 45 C) 32 to 113 F (0 to 45 C) -4 to 113 F (-20 to 45 C) Battery pack Removable battery pack with rechargeable NiCad battery, 4.8 V x 2.0 Ah Size 5.6 x 3 x 2.3 inches (14.2 x 7.6 x 5.8 centimeters) Weight 22 ounces (624 grams) Intrinsically-safe UL and cUL listed Case Impact-resistant RFI/EMI- shielded CE-approved
|